Skip to main content

Table 1 Velocities in the aneurysm and wall shear stress around the aneurysm neck for the F-LVIS and R-LVIS models

From: Efficient simulation of a low-profile visualized intraluminal support device: a novel fast virtual stenting technique

Variables

F-LVIS

R-LVIS

p Value

Velocity, m/s

 Maximum

0.14(0.09–0.26)

0.12(0.08–0.26)

0.42

 Minimum

1.44 × 10−5(4.54 × 10−6-2.86 × 10− 5)

9.72 × 10− 6(3.68 × 10− 6-2.95 × 10–5)

0.42

WSS, pa

 Maximum

12.35(9.54–26.46)

12.10(8.60–23.35)

0.06

 Minimum

2.69 × 10−3(1.04 × 10− 3-1.17 × 10−2)

3.14 × 10− 3(9.48 × 10−4-4.87 × 10− 3)

0.25

  1. F-LVIS: fast virtual stenting technique with LVIS; R-LVIS: real low-profile visualized intraluminal support; WSS: wall shear stress