From: New US device versus imaging US to assess tumor-in-brain
Experiment no. and type of US imaging system | Experiment type | US IMAGING SYSTEM | NUD | Difference Δ | |||
---|---|---|---|---|---|---|---|
Av. thickness | [mm] | ||||||
[mm] | |||||||
Muscle | Brain | Muscle | Brain | Muscle | Brain | ||
1 - GE VIVIDI | Muscle on the brain | 6.3 | 8.2 | 7.7 | 7.2 | 1.4 | 1 |
2 - BK Medical | Muscle on the brain | 8.7 | 7.1 | 7.6 | 7.2 | 1.1 | 0.1 |
3 - Sonowand | Muscle on the brain | 10.4 | 6.5 | 9.6 | 5.8 | 0.8 | 0.7 |
4 -BK Medical | Brain on the muscle | 8.2 | 8.5 | 8.1 | 8.3 | 0.1 | 0.2 |
5 - GE VIVIDI | Brain on the muscle | 12.8 | 12.2 | 13.8 | 10.2 | 1 | 2 |