From: Evaluation of the EMBOPIPE flow diverter device: in vivo and in vitro experiments
Follow-up time | Neck (mm) | Wide (mm) | Length (mm) |
---|---|---|---|
28 days (n = 6) | 1.98 ± 0.45 | 2.36 ± 0.20 | 4.68 ± 0.80 |
90 days (n = 6) | 2.03 ± 0.33 | 2.71 ± 0.39 | 4.76 ± 0.50 |
180 days (n = 7) | 2.36 ± 0.48 | 2.77 ± 0.28 | 4.89 ± 0.44 |
P value* | 0.230 | 0.064 | 0.815 |